Instruction ManualUV RadiometersUVS series
10 The formula for Rmax is: Rmax = ([VT+] - 6 V) / 1.2 A where [VT+] is the supply voltage and Rmax is the sum of the total wire resistance and
11 Data logger input channels: To prevent earth loops that influence the quality of the data from the radiometer it is recommended to use floating
12 4 CALIBRATION AND UVIATOR SOFTWARE From late 2007 the specially developed Kipp & Zonen UVIATOR Software is included with all new UVS radiom
13 The UVS Series is suitable for the measurement of UV irradiance according to theoretically defined UV-A, B and E spectra. In general all broad-b
14 Calibration Step B: The irradiances have to be corrected for the spectral mismatch error with “conversion factors”, denoted as γ (no units). Th
15 They are exposed to spectral irradiances between 280nm and 400nm (step increments 1 nm, slit width 2 nm at FWHM). The spectral measurements are
16 4.1.2 CALIBRATION STEP B: DETERMINATION OF THE CONVERSION FACTOR TABLE Without any measurement correction, a broad-band UV radiometer can
17 The solar zenith angles, Θ0, are varied between 0° and 85° (using steps of 5°) and the Ozone column densities, [O3] are varied between 200 Dob
18 Finally, the UVIATOR program corrects the UVS measurements using the appropriate solar zenith angles and Ozone column densities and makes a new
19 5 MAINTENANCE AND RECALIBRATION The quartz dome should be cleaned regularly. You may use a mild window cleansing agent which must be generously
2 IMPORTANT USER INFORMATION Reading this entire manual is recommended for full understanding of the use of this product. Should you have any c
20 6 PART NUMBERS, OPTIONS AND SPARES PARTS UVS RADIOMETERS Part No. UV-S-A-T UV-A 0354920 UV-S-B-T UV-B 0354925 UV-S-E-T U
21 APPENDIX I: ERYTHEMAL ACTION SPECTRUM ACCORDING TO CIE 1987 (DIN 5050) λ[nm] Weighting λ[nm] Weighting λ[nm] Weighting 290 1.000E+00 327 0.188E
22 APPENDIX II: CONVERSION OF OUTPUT VOLTAGE FOR INTERNAL TEMPERATURE Relation between the voltage at the temperature output (connector pin 4, yel
23 APPENDIX III: RECALIBRATION SERVICE Solar Radiation Measurement Instruments Kipp & Zonen solar radiation measurement instruments comply
24 RECALIBRATION FORM NAME : COMPANY / INSTITUTE : ADDRESS : POST CODE + CITY : COUNTRY : TEL : FAX : E-MAIL : I would like to receiv
Our customer support remains at your disposal for any maintenance or repair, calibration, supplies and spares.Für Servicearbeiten und Kali
3 DECLARATION OF CONFORMITY According to EC guideline 89/336/EEC 73/23/EEC We Kipp & Zonen B.V. Of Delftechpark 36 2628 XH Delft
4 TABLE OF CONTENTS IMPORTANT USER INFORMATION 2 DECLARATION OF CONFORMITY 3 TABLE OF CONTENTS 4 1. GENERAL INFORMATION 5 1.1 INTRODUCT
5 1. GENERAL INFORMATION 1.1 INTRODUCTION The radiometers of the UVS Series (UVS-A-T, UVS-B-T, UVS-E-T, UV-S-AB-T and UVS-AE-T) are designed fo
6 The dual band radiometers UVS-AB-T and UVS-AE-T have two separate outputs, one for the UV-A band irradiance and one for UV-B band irradiance (UV-
7 2 TECHNICAL DATA UVS-A-T UVS-B-T UVS-E-T UVS-AB/AE-TOptical Type Single band Dual band UV irradiance measured UV-A UV-B Erythemally Active
8 3 INSTALLATION When installing the radiometer you must consider: 1. The radiometer should be installed as high as possible to minimize obscurat
9 3.1 PIN CONNECTIONS OF ALL UV VERSIONS Pin connection scheme of the connector [color of wire in yellow connection cable]: 1 - V+: positive