102
Linearity ±5% from 0 to 1400 W m
-2
Response time 1 second (1/e signal)
Mechanical vibration tested up to 20 g’s without damage
Calibration reference Eppley primary standard group of pyrheliometers
Size 11 inches long
Weight 5 pounds
B 3.4 Kipp and Zonen Delft BV CH1
The pyrheliometer CH1 is designed to m easure direct solar irradiance at norm al incidence. Its m ain
characteristics are:
Built in accordance with ISO 9060 specifications for first class pyrheliometer
Slope and opening angle according to WMO recommendations
Equipped with an easily serviceable drying cartridge
Suitable for continuous outdoor use
According to clients’ specifications: (optional) internal filter, sensor temperature measurement,
extended cable and connector
Specifications
Response time 95% 7 s
99% 10 s
Zero offset: Caused by 5 K/H
change in ambient temperature ± 3 W m
-2
Non-stability < ± 1 % per year
Non-linearity ± 0.2 % (< 1000 W/m)
Spectral selectivity within
0.35 to 1.5 :m. ± 0.5 %
Temperature response (%) ± 1 %, -20 to +50
deviation due to ambient
temperature(relative to 20 °C) ± 1.5 %, -40 to +70
Tilt response None
Traceability To WRR
Sensitivity 7-15 :V W m
-1 2
Spectral range 0.2 to 4 :m, 50 % points
Impedance 50-70 S.
Irradiance 0-4000 W m
-2
Operating temperature -30 to +60 °C