Kippzonen COMBILOG Data Logger Instrukcja Użytkownika Strona 44

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 173
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 43
If the temperature is measured by external cold junction com-
pensation, a second thermocouple of the same type is re-
quired, which is connected in series with the first one. The po-
larity is selected so that the thermoelectric voltages subtract
each other. The second thermocouple is set to a fixed refer-
ence temperature ϑ
r
(mostly ϑ
r
= 0°C). The data logger then
calculates the temperature at the measuring position by means
of the linearisation curve as ϑ
x
= Lin(U
x
+Lin
-1
ϑ
r
). The data log-
ger will be informed about the reference temperature ϑ
r
via the
configuration software (“cold junction temperature”).
Hardware Manual COMBILOG 1022
44
Przeglądanie stron 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 172 173

Komentarze do niniejszej Instrukcji

Brak uwag