CFR manual
24
Appendix I: Calibration Form for pyranometers
Test pyranometer model :
………………………..
Serial number :
………………………..
Reference pyranometer model :
………………………..
Serial number :
………………………..
Sensitivity :
………………………..
V/Wm
-2
(SR)
Calibration performed by :
………………………..
Date :
………………………..
Signature :
………………………..
Reference pyranometer
Test pyranometer
Position 1
Illuminated
R :
………………..
T :
………………..
Darkened
RE :
………………..
TE :
………………..
R1 = R-RE :
………………..
T1 = T-TE :
………………..
Position 2
Illuminated
R :
………………..
T :
………………..
Darkened
RE :
………………..
TE :
………………..
R2 = R RE :
………………..
T2 = T TE :
………………..
Test pyranometer sensitivity
)21(
)21(
RR
TT
SRST
+
+
=
:
………………..
V/Wm
-2
Lamp stability
0.98 <
)22(
)11(
TR
TR
< 1.02 :
………………..
OK/Rejected